TY - JOUR
T1 - Statistical analysis of VEXLPE wear against alumina produced by a new 200-station, multidirectional pin-on-disk device
AU - Saikko, Vesa
N1 - Publisher Copyright:
© IMechE 2024.
PY - 2024/4
Y1 - 2024/4
N2 - With low wear rates shown by contemporary bearing materials of total hip prostheses, the standard deviation of wear rate is relatively high. Therefore, large sample sizes are needed for an adequate power of test. Because wear tests take a long time, it is practical to test several samples simultaneously. A new high-capacity, multidirectional wear test device, called the SuperCTPOD-200, was introduced. A 3 million-cycle wear test with an unprecedented sample size of 200 was performed for VEXLPE. The duration of the test was 6 weeks. The wear factor was normally distributed with a mean ± SD of 1.64 × 10−7 mm3/Nm ± 0.22 × 10−7 mm3/Nm (n = 200). The observation that SD was 13.1% of the mean can be useful in power analyses of future tests with other highly cross-linked polyethylenes. Burnishing was the most typical feature on the worn pins, which was in agreement with clinical findings on retrieved acetabular liners. The present study emphasizes statistics that often plays a minor role only in wear studies.
AB - With low wear rates shown by contemporary bearing materials of total hip prostheses, the standard deviation of wear rate is relatively high. Therefore, large sample sizes are needed for an adequate power of test. Because wear tests take a long time, it is practical to test several samples simultaneously. A new high-capacity, multidirectional wear test device, called the SuperCTPOD-200, was introduced. A 3 million-cycle wear test with an unprecedented sample size of 200 was performed for VEXLPE. The duration of the test was 6 weeks. The wear factor was normally distributed with a mean ± SD of 1.64 × 10−7 mm3/Nm ± 0.22 × 10−7 mm3/Nm (n = 200). The observation that SD was 13.1% of the mean can be useful in power analyses of future tests with other highly cross-linked polyethylenes. Burnishing was the most typical feature on the worn pins, which was in agreement with clinical findings on retrieved acetabular liners. The present study emphasizes statistics that often plays a minor role only in wear studies.
KW - bio-materials
KW - biotribology
KW - distribution of wear data
KW - high-capacity testing
KW - Multidirectional pin-on-disk
KW - VEXLPE
UR - http://www.scopus.com/inward/record.url?scp=85186933397&partnerID=8YFLogxK
U2 - 10.1177/09544119241235789
DO - 10.1177/09544119241235789
M3 - Article
AN - SCOPUS:85186933397
SN - 0954-4119
VL - 238
SP - 438
EP - 443
JO - Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
JF - Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
IS - 4
ER -