Preliminary results of non-contact THz imaging of cornea

Shijun Sung, James Garritano, Neha Bajwa, Sophie Deng, Jean Pierre Hubschman, Warren S. Grundfest, Zachary D. Taylor*

*Tämän työn vastaava kirjoittaja

Tutkimustuotos: LehtiartikkeliConference articleScientificvertaisarvioitu

9 Sitaatiot (Scopus)
31 Lataukset (Pure)

Abstrakti

This paper presents a novel THz optical design that allows the acquisition of THz reflectivity maps of in vivo cornea without the need for a field flattening window and preliminary imaging results of in vivo rabbit cornea. The system's intended use is to sense small changes in corneal tissue water content (CTWC) that can be precursors for a host of diseases and pathologies. Unique beam optics allows the scanning of a curved surface at normal incidence while keeping the source detector and target stationary. Basic system design principles are discussed and image sets of spherical calibration targets and corneal phantom models are presented. The presented design will enable, for the first time, non-contact THz imaging of animal and human cornea.

AlkuperäiskieliEnglanti
Artikkeli93620C
JulkaisuProceedings of SPIE - The International Society for Optical Engineering
Vuosikerta9362
NumeroJanuary
DOI - pysyväislinkit
TilaJulkaistu - 1 tammikuuta 2015
OKM-julkaisutyyppiA4 Artikkeli konferenssijulkaisuussa

Sormenjälki

Sukella tutkimusaiheisiin 'Preliminary results of non-contact THz imaging of cornea'. Ne muodostavat yhdessä ainutlaatuisen sormenjäljen.

Siteeraa tätä