NMR with excitation modulated by Frank sequences

Bernhard Bluemich*, Qingxia Gong, Eimear Byrne, Marcus Greferath

*Tämän työn vastaava kirjoittaja

Tutkimustuotos: LehtiartikkeliArticleScientificvertaisarvioitu

23 Sitaatiot (Scopus)

Abstrakti

Miniaturized NMR is of growing importance in bio-, chemical, and -material sciences. Other than the magnet, bulky components are the radio-frequency power amplifier and the power supply or battery pack. We show that constant flip-angle excitation with phase modulation following a particular type of polyphase perfect sequences results in low peak excitation power at high response peak power. It has ideal power distribution in both the time domain and the frequency domain. A savings in peak excitation power of six orders of magnitude has been realized compared to conventionally pulsed excitation. Among others, the excitation promises to be of use for button-cell operated miniature NMR devices as well as for complying with specific-absorption-rate regulations in high-field medical imaging. (C) 2009 Elsevier Inc. All rights reserved.

AlkuperäiskieliEnglanti
Sivut18-24
Sivumäärä7
JulkaisuJOURNAL OF MAGNETIC RESONANCE
Vuosikerta199
Numero1
DOI - pysyväislinkit
TilaJulkaistu - heinäk. 2009
OKM-julkaisutyyppiA1 Alkuperäisartikkeli tieteellisessä aikakauslehdessä

Sormenjälki

Sukella tutkimusaiheisiin 'NMR with excitation modulated by Frank sequences'. Ne muodostavat yhdessä ainutlaatuisen sormenjäljen.

Siteeraa tätä