Measurements of Low Optical Power with Cryostat-Based Predictable Quantum Efficient Detector at Liquid Nitrogen Temperature

Tutkimustuotos: LehtiartikkeliConference articleScientificvertaisarvioitu

11 Lataukset (Pure)

Abstrakti

We have validated optical power measurements with a Predictable Quantum Efficient Detector (PQED) at liquid nitrogen temperature (77 K) at low optical power from 130 fW to 3.3 pW. Two laser wavelengths at 514 nm and 785 nm were used. The lowest measured optical power corresponds to a photon flux of 0.5·106 photons per second (785 nm). The PQED’s responsivity is linear within the relative measurement uncertainties of 8% at 0.5·106 ph/s and 1.4% at 10·106 ph/s (95% confidence level), which enables the calibration of other low photon flux detectors directly against a primary standard of optical power.

AlkuperäiskieliEnglanti
Artikkeli012005
JulkaisuJournal of Physics: Conference Series
Vuosikerta2864
Numero1
DOI - pysyväislinkit
TilaJulkaistu - 2024
OKM-julkaisutyyppiA4 Artikkeli konferenssijulkaisussa
TapahtumaInternational Conference on New Developments and Applications in Optical Radiometry - National Physical Laboratory, Teddington, Iso-Britannia
Kesto: 11 syysk. 202315 syysk. 2023
Konferenssinumero: 15