Measurement setup for differential spectral responsivity of solar cells

Petri Kärhä*, Hans Baumgartner, Janne Askola, Kasperi Kylmänen, Benjamin Oksanen, Kinza Maham, Vo Huynh, Erkki Ikonen

*Tämän työn vastaava kirjoittaja

Tutkimustuotos: LehtiartikkeliConference articleScientificvertaisarvioitu

32 Lataukset (Pure)

Abstrakti

We have developed a setup for measuring differential spectral responsivities of unifacial and bifacial solar cells under bias light conditions. The setup uses 30 high-brightness LEDs for generating a quasi-monochromatic light source covering the wavelength range 290–1300 nm. Halogen lamps are used to generate bias-lighting conditions up to the irradiance level of 1000 W/m2. The setup has been fully characterized for spectral irradiances and spatial uniformities of all light sources. Validation measurements carried out using a reference cell of 2 × 2 cm2 area from Fraunhofer ISE demonstrated an agreement better than 2% over the wavelength range of 425–940 nm, with an expanded uncertainty of 2.6%. In the UV and IR regions, the discrepancies are higher but still within estimated uncertainties. The setup is also tested in measuring spectral responsivities of large 15 × 15 cm2 bifacial solar cells. The associated problems are discussed.

AlkuperäiskieliEnglanti
JulkaisuOPTICAL REVIEW
DOI - pysyväislinkit
TilaJulkaistu - 1 tammikuuta 2020
OKM-julkaisutyyppiA4 Artikkeli konferenssijulkaisuussa
TapahtumaJapan-Finland Joint Symposium on Optics in Engineering - Aalto University, Espoo, Suomi
Kesto: 26 elokuuta 201930 elokuuta 2019
Konferenssinumero: 13
http://oie2019.aalto.fi/

Sormenjälki Sukella tutkimusaiheisiin 'Measurement setup for differential spectral responsivity of solar cells'. Ne muodostavat yhdessä ainutlaatuisen sormenjäljen.

  • Projektit

    Advanced PV Energy Rating

    Manoocheri, F., Baumgartner, H., Kärhä, P. & Maham, K.

    01/05/201731/03/2021

    Projekti: EU other competitive funding (ERA-NET, EURAMET, EUREKA, EU JU)

    Siteeraa tätä