Abstrakti
An optical imaging system for milliKelvin temperatures has been developed based on a regular B/W surveillance camera (25 frames/s), with its CCD sensor inside the 4-K vacuum can of our nuclear demagnetization cryostat. The heat leak to the nuclear stage, caused by the operation of the video camera, was reduced below 1 nW by careful rf shielding. The construction of the system and its limits of operation are discussed.
Alkuperäiskieli | Englanti |
---|---|
Sivut | 1784-1785 |
Sivumäärä | 2 |
Julkaisu | Review of Scientific Instruments |
Vuosikerta | 65 |
Numero | 5 |
DOI - pysyväislinkit | |
Tila | Julkaistu - 1994 |
OKM-julkaisutyyppi | A1 Julkaistu artikkeli, soviteltu |
Tutkimusalat
- charge coupled device
- optical investigation changes
- superfluid He3