Calibration of RADMON radiation monitor onboard Aalto-1 CubeSat

Tutkimustuotos: Lehtiartikkelivertaisarvioitu

Tutkijat

  • Philipp Oleynik
  • Rami Vainio
  • Arttu Punkkinen
  • Oleksiy Dudnik
  • Jan Gieseler
  • Hannu Pekka Hedman
  • Heli Hietala
  • Edward Hæggström
  • Petri Niemelä
  • Juhani Peltonen
  • Jaan Praks
  • Risto Punkkinen
  • Tero Säntti
  • Eino Valtonen

Organisaatiot

  • University of Turku
  • NASU - Institute of Radio Astronomy
  • University of Helsinki

Kuvaus

RADMON is a small radiation monitor designed and assembled by students of University of Turku and University of Helsinki. It is flown on-board Aalto-1, a 3-unit CubeSat in low Earth orbit at about 500 km altitude. The detector unit of the instrument consists of two detectors, a Si solid-state detector and a CsI(Tl) scintillator, and utilizes the ΔE-E technique to determine the total energy and species of each particle hitting the detector. We present the results of the on-ground and in-flight calibration campaigns of the instrument, as well as the characterization of its response through extensive simulations within the Geant4 framework. The overall energy calibration margin achieved is about 5%. The full instrument response to protons and electrons is presented and the issue of proton contamination of the electron channels is quantified and discussed.