A Flexible and Modular Receiver Coil Array for Magnetic Resonance Imaging

Tutkimustuotos: Lehtiartikkelivertaisarvioitu

Tutkijat

Organisaatiot

  • National Taiwan University

Kuvaus

We propose a flexible form-fitting MRI receiver coil array assembled by individual coil modules. This design targets MRI applications requiring a receiver array conforming to the anatomy of various shapes or sizes. Coil modules in our proposed array were arranged with gaps between them. Each coil module had a circumferential shielding structure stacked on top of the coil. Together they achieve robust decoupling when the array was bent differently. Two types of the circumferential shielding structure were investigated by using full-wave electromagnetic simulations and imaging experiments. Results showed that our flexible coil array had good decoupling between coils whether they were on a flat or curved surface with the S21 magnitude ranged between -18.1 dB and -19.9 dB in simulations, and with the average of off-diagonal entries of the noise correlation matrix less than 0.047 in experimental measurements. Anatomical images of human brain, calf, and knee were acquired by our seven-channel prototype on a 3T MRI system. The maximal and the average SNR within 50 mm from our array surpassed those from the commercial 32-channel head and 4-channel flexible coil arrays by 2.63/1.35-fold and 3.89/1.50-fold, respectively.

Yksityiskohdat

AlkuperäiskieliEnglanti
Artikkeli8478795
Sivut824-833
Sivumäärä10
JulkaisuIEEE Transactions on Medical Imaging
Vuosikerta38
Numero3
Varhainen verkossa julkaisun päivämäärä1 tammikuuta 2018
TilaJulkaistu - maaliskuuta 2019
OKM-julkaisutyyppiA1 Julkaistu artikkeli, soviteltu

ID: 28771046