TY - JOUR
T1 - A Comparison Between a Resuscitation Glove and Standard Manual Compressions on the Quality of Cardiovascular Resuscitation Manikin-Based Randomized Crossover Trial
AU - Kahsay, Desale Tewelde
AU - Tommila, Miretta
AU - Peltonen, Laura Maria
AU - Löyttyniemi, Eliisa
AU - Xiao, Yu
AU - Mauranen, Henry
AU - Salanterä, Sanna
N1 - Publisher Copyright:
Copyright © 2025 The Author(s).
PY - 2025
Y1 - 2025
N2 - Background: Several audiovisual feedback (AVF) devices have been developed to monitor chest compression quality during cardiopulmonary resuscitation (CPR). However, most marketed stand-alone AVF devices are inflexible and rigid, causing discomfort and sometimes pain to the rescuers' hands. Objective: The objective of this study was to evaluate the effectiveness and usability of a newly developed soft and flexible resuscitation glove designed to improve the quality of chest compressions during CPR. Methods: We conducted a manikin-based randomized crossover study to compare the effectiveness of a newly developed AVF device (ResuGlove CPR Group) and standard CPR (Standard CPR Group) in improving the quality of chest compressions in simulated cardiac arrest cases. The usability of the newly developed ResuGlove was assessed using a System Usability Scale questionnaire. Results: There were no significant differences in compression depth (mean, 53.69 vs 53.28; P =.70) and compression rate (mean, 111.48 vs 113.38; P =.23) between the ResuGlove CPR and Standard CPR groups. However, the group using ResuGlove had a higher percentage of complete chest releases between compressions (P =.008). Furthermore, the ResuGlove CPR Group had a significantly higher percentage of participants who performed chest compressions with adequate compression depth (82.8% vs 41.4%, P =.001) and compression rate (96.6% vs 72.4%, P =.012) compared with the Standard CPR Group. The ResuGlove usability score was calculated to be 70.4. Conclusions: The newly developed ResuGlove significantly improved the quality of certain chest compression parameters, and the device's usability score was within the acceptable range.
AB - Background: Several audiovisual feedback (AVF) devices have been developed to monitor chest compression quality during cardiopulmonary resuscitation (CPR). However, most marketed stand-alone AVF devices are inflexible and rigid, causing discomfort and sometimes pain to the rescuers' hands. Objective: The objective of this study was to evaluate the effectiveness and usability of a newly developed soft and flexible resuscitation glove designed to improve the quality of chest compressions during CPR. Methods: We conducted a manikin-based randomized crossover study to compare the effectiveness of a newly developed AVF device (ResuGlove CPR Group) and standard CPR (Standard CPR Group) in improving the quality of chest compressions in simulated cardiac arrest cases. The usability of the newly developed ResuGlove was assessed using a System Usability Scale questionnaire. Results: There were no significant differences in compression depth (mean, 53.69 vs 53.28; P =.70) and compression rate (mean, 111.48 vs 113.38; P =.23) between the ResuGlove CPR and Standard CPR groups. However, the group using ResuGlove had a higher percentage of complete chest releases between compressions (P =.008). Furthermore, the ResuGlove CPR Group had a significantly higher percentage of participants who performed chest compressions with adequate compression depth (82.8% vs 41.4%, P =.001) and compression rate (96.6% vs 72.4%, P =.012) compared with the Standard CPR Group. The ResuGlove usability score was calculated to be 70.4. Conclusions: The newly developed ResuGlove significantly improved the quality of certain chest compression parameters, and the device's usability score was within the acceptable range.
KW - cardiac arrest
KW - cardiopulmonary resuscitation
KW - feedback
KW - medical device