Needle bevel geometry influences the flexural deflection magnitude in ultrasound-enhanced fine-needle biopsy

Saif Bunni, Heikki J. Nieminen*

*Corresponding author for this work

Research output: Contribution to journalArticleScientificpeer-review

1 Citation (Scopus)
92 Downloads (Pure)

Abstract

It has been recently demonstrated that use of ultrasound increases the tissue yield in ultrasound-enhanced fine-needle aspiration biopsy (USeFNAB) as compared to conventional fine-needle aspiration biopsy (FNAB). To date, the association between bevel geometry and needle tip action has not been widely explored. In this study, we studied the needle resonance characteristics and deflection magnitude of various needle bevel geometries with varying bevel lengths. With a conventional lancet, having a 3.9 mm long bevel, the tip deflection-to-power ratio (DPR) in air and water was 220 and 105 µm/W, respectively. This was higher in comparison to an axi-symmetric tip, having a bevel length of 4 mm, which achieved a DPR of 180 and 80 µm/W in air and water, respectively. This study emphasised the importance of relationship between flexural stiffness of bevel geometry in the context of various insertion media and, thus, could provide understanding on approaches to control post-puncture cutting action by modifying the needle bevel geometry, essential for the USeFNAB application.

Original languageEnglish
Article number17096
JournalScientific Reports
Volume12
Issue number1
DOIs
Publication statusPublished - 12 Oct 2022
MoE publication typeA1 Journal article-refereed

Fingerprint

Dive into the research topics of 'Needle bevel geometry influences the flexural deflection magnitude in ultrasound-enhanced fine-needle biopsy'. Together they form a unique fingerprint.

Cite this