High frequency circular translation pin-on-disc wear testing of UHMWPE using a ball-on-flat contact along a hypotrochoidal track

Research output: Contribution to journalArticleScientificpeer-review

Researchers

Research units

Abstract

To intensify experimental research within the field of orthopaedic tribology, a three-station, dual motion, high frequency (25.3 Hz) circular translation pin-on-disc wear test device was recently introduced. In the present study, the pins were CoCr with a spherical, polished bearing surface of 28 mm radius, whereas the flat discs were conventional UHMWPE. This configuration was intended to simulate the wear mechanisms of total knee prostheses. The number of wear cycles run was as high as 200 million. The mean wear rate was 0.35 mg per one million cycles (0.77 mg/24 h) which corresponded to a mean wear factor of 3.5 × 10−6 mm3/Nm. The study provided further proof that a wear test for orthopaedic implant materials can be accelerated by substantially increasing the cycle frequency, provided that the sliding velocity remains close to the values obtained from biomechanical studies. Hence, the moderate frictional heating will not lead to unrealistic wear mechanisms.

Details

Original languageEnglish
Pages (from-to)149-152
Number of pages4
JournalPolymer Testing
Volume60
Publication statusPublished - 1 Jul 2017
MoE publication typeA1 Journal article-refereed

    Research areas

  • Accelerated wear testing, Biotribology, Orthopaedic tribology, Prosthetic joints, UHMWPE