Effect of serum dilution fluids on the wear of unirradiated and high dose gamma-irradiated, vitamin E stabilized UHMWPE

Research output: Contribution to journalArticleScientificpeer-review

Researchers

Research units

Abstract

In orthopaedic tribology, the ISO 14242-1 standard recommends that the serum lubricant is diluted with deionized water (DW). Sometimes, phosphate buffered saline (PBS) is used instead for the dilution. Noncyclic, multidirectional pin-on-disk tests showed that the dilution of alpha calf serum with PBS or phosphate buffer resulted in a 5–40-fold decrease in the wear factor of unirradiated and of vitamin E stabilized, highly crosslinked ultrahigh molecular weight polyethylene (UHMWPE) compared with DW dilution. It appears that the phosphates were mainly responsible for this phenomenon. PBS dilution may result in a substantial underestimation of clinical wear. With 1:1 DW dilution, the mean wear factor of highly crosslinked UHMWPE was 5.7-fold lower than that of unirradiated UHMWPE.

Details

Original languageEnglish
Pages (from-to)76-80
Number of pages5
JournalWear
Volume430-431
Publication statusPublished - 15 Jul 2019
MoE publication typeA1 Journal article-refereed

    Research areas

  • Biomedical devices, Lubrication environment, Noncyclic pin-on-disk, UHMWPE