An orthogonal shim coil for 3T brain imaging

Jiazheng Zhou, Jason P. Stockmann, Nicolas Arango, Thomas Witzel, Klaus Scheffler, Lawrence L. Wald, Fa-Hsuan Lin*

*Corresponding author for this work

Research output: Contribution to journalArticleScientificpeer-review

13 Citations (Scopus)

Abstract

Purpose: We designed and implemented an orthogonal shim array consisting of shim coils with their planes perpendicular to the planes of neighboring RF coils. This shim coil improves the magnetic field homogeneity by minimizing the interference to RF coils. Methods: Using realistic off-resonance maps of the human brain, we first evaluated the performance of shim coils in different orientations. Based on simulations, we developed a 7-channel orthogonal shim array, whose coil plan was perpendicular to neighboring RF coils, at the forehead. A programmable open-source current driver supplied shim currents. Results: The 7-channel orthogonal shim array caused only marginal SNR loss to the integrated 32-channel RF-shim array. The 7-channel orthogonal shim array itself improved the magnetic field homogeneity by 30% in slice-optimized shimming, comparable to the baseline shimming offered by the scanner's 2nd order spherical harmonic shimming. Conclusion: Orthogonal shim coils can improve the field homogeneity while maintaining high image SNR.

Original languageEnglish
Pages (from-to)1499-1511
Number of pages13
JournalMagnetic Resonance in Medicine
Volume83
Issue number4
Early online date21 Oct 2019
DOIs
Publication statusPublished - 1 Apr 2020
MoE publication typeA1 Journal article-refereed

Keywords

  • B-0 homogeneity
  • multi-coil
  • shimming
  • INTEGRATED PARALLEL RECEPTION
  • MAGNETIC-FIELD HOMOGENEITY
  • OPTIMIZATION
  • EXCITATION
  • ROBUST

Fingerprint

Dive into the research topics of 'An orthogonal shim coil for 3T brain imaging'. Together they form a unique fingerprint.

Cite this